    EchoPod® UG06 & UG12 Reflective Ultrasonic Liquid Level Transmitter

    DOWNLOADS

     

     

    Application

    The general purpose reflective ultrasonic level transmitter provides continuous level measurement up to 39.3’ (12m) with a 4-20 mA analog signal output, and is configured via its integral push button display module or WebCal software. The non-contact liquid level sensor features our proprietary Reflective Technology™ that delivers reliable level measurement in condensing environments. Select this sensor for bulk tanks with non-foaming or highly vaporous media such as chemicals, water, wastewater and oil. Typical applications include bulk storage, neutralization tank, clarifier and waste sumps.

    Features

    Offered in 6m (19.6’) and 12m (39.3’) measurement ranges

    Reflective Technology™ measures reliably with condensation

    Corrosion resistant PVDF transducer with IP68 PP enclosure

    Fail-safe diagnostics with selectable signal fail-safe outputs

    LCD display indicates level in inches, meters or percent of span

    Narrow 3” (7.6cm) beam width for applications with limited space

    Windowed enclosure cap provides liquid tight level indication

    Configuration via push button display or WebCal software

    Automatic temperature compensation from -40º to 80º C

    Reflective

    Condensation is the most common variable in liquid level applications. Condensation attenuates the acoustic signal of ultrasonic sensors with horizontal transducers, weakening their signal strength and signal to noise ratio by up to 50%, and substantially reducing their measurement reliability. At the core of Reflective Technology™ is a simple fact. Unlike flat horizontal surfaces, significant water droplets cannot adhere to smooth vertical surfaces. By orienting the transducer vertically, condensation runs off the transducer and does not affect sensor performance. The unimpeded transmit and receive signals are redirected to and from the liquid off a 45º reflector, delivering reliable level measurement. Thanks gravity.

    Reflective Transducer

    Patent Pending

    Reliable Measurement

    Horizontal Transducer

    Signal Attenuation

     

     

     

    Specifications

    Range: UG06: 8” to 19.6’ (20cm to 6m), UG12: 18” to 39.3’ (45.7cm to 12m)
    Accuracy: ± 0.2% of range
    Resolution: UG06: 0.079” (2mm), UG12: 0.196” (5mm)
    Dead band: UG06: 8” (20.3cm), UG12: 18” (45.7cm)
    Beam width: UG06: 3” (7.6cm), UG12: 6” (15.2cm)
    Configuration: Push button or WebCal® PC Windows® USB® 2.0
    Memory: Non-volatile
    Display type: LCD, 6-digit
    Display units: Inch, cm or percent
    Supply voltage: 14-28 VDC
    Max. consumption: 0.5W
    Loop resistance: 500 ohms @ 24 VDC
    Signal output:

    4-20mA, two-wire

    Signal invert: 4-20mA or 20-4mA
    Signal fail-safe: 4mA, 20mA, 21mA, 22mA, hold last
    Process temp.:

    F: -40º to 176º

    C: -40º to 80º

    Temp. comp.:

    Automatic

    Ambient temp.:

    F: -31º to 140º

    C: -35º to 60º

    Pressure: 30 psi (2 bar)
    Enclosure rating: Type 6P (IP68)
    Encl. material: Polypropylene
    Encl. cap material: Clear polycarbonate
    Enclosure vent: Water tight membrane
    Conduit entrance: 1/2” NPT
    Transducer type: Reflective
    Transducer mat.: Polyvinylidene fluoride
    Process mount: -0001: 2” NPT, -0011: 2” G
    Mount gasket: -0001: N/a, -0011: Viton®
    Classification: General purpose
    Approval (pending): UL 61010-1
    Compliance: CE, RoHS