SI100 & SI200 - Specification
Orbit® ACS Digital Gauging Probes
Product
|
|
|
|
|
|
|
Spring Push Axial Cable
|
SIxxx P/1/S
|
SIxxxP/2/S/2
|
SIxxxP/5/S
|
SIxxxP/10/S
|
SIxxxP/20/S
|
SIxxx6P/2/S
|
Spring Push Radial Cable
|
SIxxxPR/1/S
|
SIxxxPR/2/S
|
SIxxxPR/5/S
|
SIxxxPR/10/S
|
SIxxxPR/20/S
|
N/A
|
Pneumatic Axial Cable
|
N/A
|
SIxxxP/2/P
|
SIxxxP/5/P
|
SIxxxP/10/P
|
SIxxxP/20/P
|
N/A
|
Pneumatic Radial Cable
|
N/A
|
SIxxxPR/2/P
|
SIxxxPR/5/P
|
SIxxxPR/10/P
|
SIxxxPR/20/P
|
N/A
|
Measuring Range (mm)
|
1
|
2
|
5
|
10
|
20
|
2
|
Body Diameter
|
|
|
Note: SIxxx can be either SI100 or SI200
|
Performance
|
Accuracy (% of Reading) (1)
|
0.2
|
0.15
|
0.2
|
Repeatability μm (2)
|
0.15
|
Resolution μm -user selectable
|
<0.01
|
<0.05
|
<0.1
|
<0.01
|
Alarm Outputs - selectable High, OK, Low
|
3 outputs either NPN, PNP, logic Programable Active Hi or Lo
|
Discrete Inputs - user selectable
|
eg. Print, Zero, Preset (see manual for other options)
|
Update Rate for I/O discrete (ms)
|
5
|
Bandwidth of Electronics (Hz) - user selectable
|
460, 230, 115, 58, 29, 14, 7, 4
|
Bandwidth of Electronics (Hz) - user selectable
|
MODBUS (RTU or ASCII) or Solartron Serial Formats
|
Communications Interface Hardware
|
RS485 or RS232 (User selectable) Up to 115,200 Baud
|
Update Rate for Serial Data (ms)
|
25
|
Pre Travel ( mm)
|
0.15
|
Post Travel (mm)
|
0.35
|
0.85
|
0.35
|
Tip Force (N) at Middle of Range ±20%
|
|
Spring Push
|
0.7
|
Pneumatic at 0.4 bar
|
N/A
|
0.8
|
0.85
|
0.7
|
N/A
|
Pneumatic at 1 bar
|
N/A
|
2.8
|
2.5
|
N/
|
|
|
IP65 with gaiter or IP50 without gaiter
|
Sealing for Probe Interface Electronics
|
Top and Front: IP41, Rear: IP20, In line connector: IP67
|
|
-20 to +70
|
Probe Operating Temperature with Gaiter (ºC)
|
+5 to +80
|
Probe Operating Temperature without Gaiter (ºC)
|
-10 to +80
|
Electronics Operating Temperature (ºC)
|
0 to 60
|
|
Emissions EN61000-6-3, EN61326
|
|
Immunity EN61000-6-2, EN61326
|
|
18 to 32 VDC @ 0.07A typical
|
|
|
Stainless Steel
|
|
Nylon, Ruby, Silicon Nitride, Tugsten Carbide
|
|
Fluoroelastomer
|
|
PUR
|
|
ABS
|
[1] Accuracy 0.1μm or % reading, whichever greater
[2] Obtained by repeated operation against a carbide target with side load applied to bearing
Instrumentation Functionality
|
Measurement Modes - SI100
|
A, MAX-MIN
|
Measurement Modes - SI200
|
A,B, A+B, A-B, (A+B)/2, (A-B)/2, MAXA-MINA, MAXB-MINB
|
|
Track, Peak+, Peak-
|
|
Absolute, Zero (tare), Preset
|
|
mm, inches or mils
|
|
|
Bar representing reading
|
|
Digital up to 5 decimal places mm (6 for inches)
|
|
Red bar and red digital reading indicates measurement outside of limits
|
|
|
|
|
Product and Technology News
Pressure and Temperature Transmitter
Download Catalogue
|