All data overview

Modul type

Loggito series
Analog inputs  
Sensors mV, mA, thermocouples, RTDs (no RTDs with channel types 8 AI and 4 AI)
Sum sampling rate 1/3 Hz .. 1 kHz
Voltage measurement range ±78 mV .. ±10 V
Current measuring range 0 .. 20 mA / 4 .. 20 mA / ±20 mA / any (with freely selectable external shunts)
Optional additional boards
(only with 24 V supply)
Software switchable internal shunt for active and passive 4 .. 20 mA sensors
Software switchable feed for passive 4 .. 20 mA sensors
Resistance measurement range 0 .. 60 kΩ
Resolution 24 Bit
Input impedance > 5GΩ
Dielectric withstand voltage
(tolerable overvoltage)
±24 VDC
Max. differential voltage channel to channel ±24 VDC
Galvanic isolation against FE ±400 VDC
Precision of current and voltage measurement 0.01% from measurement range end value
RTD precision 0,1 K
Thermocouples precision 0.3 K (external thermoblock), < 1.0 K (internal reference junction)
Analog outputs  
Resolution 12 bit
Max. output rate 1 Hz
Output range 4 .. 20 mA
Max. load 550 Ω
Galvanic isolation output to analog inputs none
Galvanic isolation to system / FE ±400 VDC
Digital / frequency inputs  
Input signal low low: 0 .. 1 V / high: 3,6 .. 100 VDC@3,5 mA
Frequency input measuring range
(combined DIOs)
0.2 Hz .. 500 Hz
Frequency input measuring range (pure DIs) 0.2 Hz .. 30 kHz
Counter width 64 bit
Galvanic isolation ±400 VDC against FE / other channels (DIOs in groups of 2 channels)
Digital / PWM outputs  
Grouping 2 per group
Max. switching voltage / current 50 V / 2.5 A
PWM basic frequency / duty cycle 20 Hz .. 10 kHz / 1:100
Galvanic isolation ±400 VDC against FE / channels of other groups
Modul type Loggito Logger Loggito Wireless Loggito USB
Interfaces  
LAN 10 x Base-T / 100 Base-Tx           -           -
WLAN 802.11b/g/n, max. 72 MBit/s
(optional, alternative to radio network interface)
          -           -
Radio network interface IEEE802.15.4, max. 256 kBit/s
(optional, alternative to WLAN)
IEEE802.15.4,
max. 256 kBit/s
          -
USB device – mini socket 2.0 low/full/high speed           - 2.0 low/full speed
USB host – type A socket 2.0 low/full speed           -           -
Protocols  
RS 232/485 (via USB extension) Modbus RTU Master, Modbus RTU slave, SCPI, ASCII           -           -
TCP/IP Modbus TCP, OPC UA (DA) client server           -           -
Data storage  
Internal 4 / 8 / 16 GB / approx. 30 million measurement values per GB           -           -
External USB, NFS, CIFS, (S)FTP           -           -
General technical information  
Dimensions L x W x H approx. 96 mm x 68 mm x 65 mm
Weight without battery approx. 250 g, with battery approx. 300 g
Fixing mounting rail DIN EN 60715 or screw mounting horizontally / vertically or upwards
Connecting terminals plug-in spring terminals, max. 24 terminals in 2 rows,
conductor cross-section (rigid / flexible) 0.2 .. 2.5 mm², ferrule 0.25 .. 1.5 mm²
Temperature range -20 .. 60° C
Humidity max. 90 % relative humidity, non-condensing
Supply voltage alternatively 12 .. 24 VDC / ±20 % or
5 VDC (via mini USB connector) or
integrated accumulator 3.7 V, 2050 mAh (optional)
12 .. 24 VDC / ±20 % or
integrated accumulator
3.7 V, 2050 mAh (optional)
5 VDC
(via mini USB connector)
Power input

max. 3 watts plus power consumption of current loop for passive 4 .. 20 mA sensors, sleep mode max. 50 mW, hibernation mode max. 1 mW

max. 1 watt max. 1 watt
 
 

 

 

Product and Technology News