LCM
How it Works
- 1. LCM series or LSB series load cells are threaded in-line with a linear actuator and syringe piston measuring the force the linear actuator is applying to the syringe piston for test validation.
- 2. As the linear actuator pushes down on the load cell/piston assembly the flow of fluid pushes open the TAVR valve.
- 3. A check valve keeps the fluid moving only through the TAVR valve and a pressure vessel is used as a reservoir allowing the system to reach the burst testing pressures.
- 4. The linear actuator then retracts, pulling the piston up, forcing the TAVR closed and refilling through the now open check valve.
- 5. The USB220 displays and logs the data to a PC via FUTEK's SENSITâ„¢ software. This data can be used to validate the accuracy and precision of the pressure generated by the syringe piston.