LSB200
How it Works
- 1. An artificial restriction (occlusion) is placed at the desired point using a clamp.
- 2. One end of the clamp has FUTEK's LSB200 mounted between the clamp and the mount point.
- 3. By creating an artificial flow restriction (an occlusion), the fluid exerts an increased force onto the tubing.
- 4. The LSB200 measures the force applied to the occlusion by the fluid which can be correlated to fluid flow rate.
- 5. The LSB200 was chosen to provide a high accuracy and linearity force measurement with significant overload protection.
- 6. FUTEK's IAA and IDA100 amplify and transmit the load cell signal to the system PLC for flow measurement.
- 7. Additionally, the IDA100 enables logging and monitoring of the load cell output via a PC with FUTEK's SENSIT software.